Melt Mass-Flow Rate Test Machine, ASTM D1238-98 FTech-ISO1133
  • Product Introduction
Introduction:
This machine is deisgned according to  ISO1133, ASTM D1238-98 ,ISO1133; Melt flow rate was measured using automatic sampling, weighing scales manner; then, according to the formula to calculate the volumetric flow rate. Final results of the measurement are displayed on the LCD screen of the instrument, micro-printer output. The instrument measurement data is accurate, simple, intuitive, stable and reliable performance; It is suitable for fluorine plastic, nylon and other engineering plastics, it is also used in polyethylene (PE), polypropylene (PP), polyacetal (POM), polyphenylene ethylene (PS) ABS resin, polycarbonate and other plastics - Determination of melting temperature lower. The instrument is easy to operate, the data is accurate, safe and reliable, widely used in the detection of the relevant departments in the field of plastic industry;
Model:FTech-ISO1133
Standard:
ASTM D1238-98, ISO1133
Features:
1.The product uses PLC control, LCD display: time, temperature, sampling times.
2. Quality Act (MFR) measurement, high precision, accurate data.
1.Test Method: Quality Act (MFR)
2. Eight standard test load
1: 0.325kg = (piston + weight + pallet weight body lagging number +1) = 3.187N
2: 1.200 kg = (0.325 + 0.875 weight 2) = 11.77 N
3: 2.160 kg = (0.325 + 1.835 weight 3) = 21.18 N
4: 3.800 kg = (0.325 + 3.475 weight 4) = 37.26 N
5: 5.000 kg = (0.325 + 5 #4.675 weight) = 49.03 N
6: 10.000 kg = (0.325 + 5 # 4.675 5.000 weight weight No. +6) = 98.07 N
7: 12.500 kg = (0.325 + 5 # 4.675 +6 weight number weight 2.500 5.000 + 7) = 122.58 N
8: 21.600 kg = (0.325 + 0.875 weight number 2 + 3 + 4 1.835 5 4.675 + 3.475 + 5.000 + 6 7 8 2.915 2.500 + weight) = 211.82 N
Remarks; test load relative error ≤0.5%.
3.Temperature range 20 ~ 400 ºC; accuracy of ± 0.2 ºC; resolution 0.1 ºC
4.The test device:
Spout diameter: Φ2.095 ± 0.005mm
Spout length: 8.000 ± 0.025mm
Loading cartridge diameter: Φ9.550 ± 0.025mm
Loading cartridge length: 152 ± 0.1mm
Piston head diameter: 9.475 ± 0.015mm
Piston head length: 6.350 ± 0.100mm
5. Display: LCD display
6. The test report print at any time, to facilitate consolidation
7. Scraper: Auto / Manual
8. The power supply voltage: 220V ± 10% 50Hz
9. The instrument Dimensions: 510 × 370 × 600
10. Weight (kg): Host 25kg weight 21.6kg
11. Temperature range: --- 400 ºC.
12. Temperature accuracy: ± 0.2 ºC
13. Resolution: 0.1 ºC
14 .Working conditions: ambient temperature of 10 ºC -40 ºC; ambient relative humidity of 30% -80%; non-corrosive surrounding media, properly installed and adjusted to the level of a solid foundation. No strong air convection; around no vibration, no strong magnetic interference.
15. Power: 0.45KW
16. The index range: 0.1-400.00g / 10min (MFR)
17. Timing accuracy: 1S
18. Load: (0.325kg; 0.875kg; 0.96kg; 1.2kg; 1.64kg; 5kg;)
19. Cut material modes: manual, time control
20. Barrel inner diameter: 9.550mm ± 0.025mm
21. Barrel length: 160mm
22 .Die: made of tungsten carbide, length: 8.000mm ± 0.025mm inner diameter: 2.095mm ± 0.005mm
Related Products